產品用途德國RSI 1000升腐蝕試驗柜
試驗箱:
-材質:特殊不銹鋼
-絕緣材料:60mm聚氨酯
-在鹽霧過程中,腔室通過底部水加熱
-在氣候控制模式下,通過由特殊鈦制成的壁式換熱器提供加熱和氣候控制
-由多層中空玻璃制成的透明試驗室門
-GRP地板格柵
-用于吹掃試驗室的壓縮空氣吹掃
-*大載荷:300kg
-包括140l鹽溶液儲罐
布
該系統(tǒng)由兩部分組成:
-右側安裝有控制面板的試驗箱
-單獨安裝的溫度控制單元
試驗室尺寸:
-寬度:1000毫米
-深度:900毫米
-高度(至屋面坡度起點):1030mm
-高度(至屋頂頂點):1350mm
外部尺寸:
-根據附圖
溫度控制和空調:
-冷凝水操作
-通過底部水供暖
-溫度范圍:室溫至+50°C
-鹽霧
-通過地板水/墻壁熱交換器供暖
-溫度范圍:室溫至+50°C
-氣候階段
-通過壁式換熱器進行溫度控制
-通過集成空氣循環(huán)系統(tǒng)進行空氣循環(huán)
-溫度范圍:+20至+60°C
-相對濕度范圍:30%至95%
-霜凍階段
-通過壁式換熱器進行溫度控制
-通過集成空氣循環(huán)系統(tǒng)進行空氣循環(huán)
-溫度范圍:室溫低至-20°C
技術參數鹽霧:
-通過2個鈦制噴嘴產生霧
-通過2個精密計量泵(隔膜泵)進行流量控制
-鹽霧沉降量可在1.5至3 ml/h之間調節(jié)
-鹽水儲罐140升
溫度控制單元:
-水冷(所需流動溫度6-25°C,所需壓差>1巴)
-制冷劑:R449a
控制系統(tǒng)(RSI Com10):
-通過10英寸觸摸屏操作
-溫度/相對濕度曲線的圖形顯示
檢測項目A full test cycle lasts 7 (Fig. 1) days and consists of:
3 times = 24h cycle A (Fig. 2; see also 7.2) without freezing phase, with 3 h salt spray phase (1 % NaCl solution)
3 times = 24h cycle B (Fig. 3; see also 7.2) without freezing phase, without salt spray, including assessment of specimens (or walk-around inspection)
1 time = 24h cycle C (Fig. 4; see also 7.2) with freezing phase, without salt spray

Fig.1: The placement of specimens in the test chamber shall be carried out exclusively during the
first cycle B of the evaluation phase. This also applies for the removal of the specimens.

Fig.: 2: 24 h-cycle A, with 3 h salt spray phase (1 % NaCl)

Fig.: 3: 24 h-cycle B without freezing phase, without salt spray, with evaluation phase

Fig. 4: 24 h-cycle C, with freezing phase, without salt spray
5.2 Test interruption
Interruption of the test or opening of the test chamber is only permitted during the evaluation phase (cycle B). The test chamber may not be switched off during this time.
No interruption of the salt spray phase is permitted. Longer interruptions of the cyclic climate phases are to be avoided; if unavoidable, they must be mentioned in the test report.
5.3 Placing specimens in the chamber
The specimens shall preferably be positioned in the chamber at an angle of 65° to 75° relative to the horizontal with the side to be tested uppermost. Components are to be positioned as agreed with the principal, preferably in the orientation in which they will be installed. The specimens (or components) may not contact each other or shield other specimens from the salt spray.
When using two separate test chambers for the salt spray phase and the cyclic climate phase, the specimens may not be allowed to dry-out when being transferred from one chamber to the other.
Vehicle components or special corrosion test plates may also be used as specimens for the test. In the case of components, special care shall be taken to ensure that salt solution does not pool on the specimen, as this will affect the test conditions (climate). If this cannot be avoided, such pooling must be removed at least once a week.
關鍵詞: